A-87.341.000
Swansensor Shurecon S
4-Electrode conductivity sensor with built-in Pt1000-temperature probe. Excellent stability and fouling resistance. For sample temperatures up to 120°C. For the connection to the Solicon4 conductivity transmitters.
Details
Sensor for the measurement of specific conductivity.
For applications in surface water, potable water and cooling water.
Four electrode design with integrated temperature sensor.
Unaffected by fouling.
No measuring errors due to polarization effects.
Sterilizable
Sensor with plug.
For the use together with SWAN conductivity transmitters AMI Solicon4 or AMU Solicon4.
| Recommended conductivity range 0.1 µS/cm ... 100 mS/cm | |
|---|---|
| Accuracy | ± 1% or ± 0.2 µS/cm; whichever is greater |
| Cell constant k | 0.45 cm-1 |
| Temperature sensor type | Pt1000, DIN class A |
| Operating conditions: | |
| max. temperature | 120°C at 6.5 bar |
| max. pressure | 12 bar at 20°C |
| sterilizable at | 120°C / 5 bar / 30min. |
| Materials: | |
| Body | PPSU |
| Shaft | stainless steel SS316L |
| Current electrodes | SS 316L |
| Isolation | PEEK |
| Voltage electrodes | Platinum |
| Cover | PVDF |
| Electrical connection | Plug M16 male (IP67) |
| Process connection | G1" thread |
| Free space around sensor tip | 20 mm |
Downloads
Products
Related Products
There are no available products for related products.